FDAOctober 19, 2017device

CombiDiagnost R90 Image-Intensified fluoroscopic x-ray system CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The tube adapter-plate green/yellow protective earthing (PE) wire was removed. and spacers with different thickness were used to position pre-loc ring in respect to focal spot in specified distance.

What to do

FDA enforcement status: Terminated

Brands named

philips electronics north americaphilipsphilips electronics