FDASeptember 25, 2023device
Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q160Z, GIF-Q180, GIF-...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
What to do
FDA enforcement status: Ongoing
Brands named
aizu olympusaizu
UPCs
049531703433600495317030749204953170339745049531703341150495317033976904953170307539049531703052900495317030527604953170312373049531703075080495317030748504953170066795
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2113.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-21122026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2118.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2110. May be included with the Olympus OER-ELITE.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-21112026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2117.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2116.2026-03-25
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →