FDASeptember 25, 2023device

Hysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

What to do

FDA enforcement status: Ongoing

Brands named

aizu olympusaizu

UPCs

04953170340178

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Hysterofiberscope-For endoscopic diagnosis within the uterus. Model Number: HYF-XP. — Recall Details · AllClear