FDASeptember 25, 2023device

Tracheal fiberscope-airway management, which includes observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: LF-DP, LF-GP, LF-TP.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

What to do

FDA enforcement status: Ongoing