Colonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). Model Numbers: PCF-H180AI, PCF-H180AL, PCF-H190DI, PCF-H190DL, PCF-H190I, PCF-H190L, PCF-H190TI, PCF-H190TL, PCF-HQ190I, PCF-HQ190L, PCF-PH190I, PCF-PH190L, PCF-...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
What to do
FDA enforcement status: Ongoing
Brands named
aizu olympusaizu
UPCs
049531703398750495317033989904953170363696049531703636720495317030521404953170305191049531704207190495317042070204953170416132049531704161180495317030525204953170305238
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2113.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-21122026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2118.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2110. May be included with the Olympus OER-ELITE.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-21112026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2117.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2116.2026-03-25
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →