FDASeptember 25, 2023device

Colonoscope-indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). Model Numbers: PCF-H180AI, PCF-H180AL, PCF-H190DI, PCF-H190DL, PCF-H190I, PCF-H190L, PCF-H190TI, PCF-H190TL, PCF-HQ190I, PCF-HQ190L, PCF-PH190I, PCF-PH190L, PCF-...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

What to do

FDA enforcement status: Ongoing

Brands named

aizu olympusaizu

UPCs

049531703398750495317033989904953170363696049531703636720495317030521404953170305191049531704207190495317042070204953170416132049531704161180495317030525204953170305238

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.