FDAOctober 13, 2015device

Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher. The I.V. poles are stationary, height adjustable stands intended to hold infusion liquids, infusion accessories, and/or other medical devices. They may be used on the following parent devices: Model...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

It was identified by a customer complaint that some I.V. poles were retracting/lowering from an extended position to the low height unexpectedly. An investigation was initiated which identified that the I.V. Latch Housing was nonconforming. potential hazards: " Suspended I.V. pole retracts unexpectedly " Suspended I.V. pole retracts unexpectedly with pump/bags attached

What to do

FDA enforcement status: Terminated

Brands named

stryker medical divisionstrykerstryker medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.