FDAOctober 25, 2022device

GREER DIA - KIT ANTIGEN M. FAENI, 2.0mL 20 MG/ML, Analyte Specific Reagent, Item: K62, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency. The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the lot was misinterpreted.

What to do

FDA enforcement status: Terminated

Brands named

greer laboratoriesgreer