FDAOctober 5, 2021device

Trinity Biotech HbA1c (GHb) Contorls Kit, 500uL (Levels I & II), REF 01-04-0020, Part number 05-01-0040B, A1c analysis

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. Additionally, per the firm, the unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.

What to do

FDA enforcement status: Terminated

Brands named

primus corporation dba trinity biotechprimusprimus corporation

UPCs

0539151674808

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Trinity Biotech HbA1c (GHb) Contorls Kit, 500uL (Levels I & II), REF 01-04-0020, Part number 05-01-0040B, A1c analysis — Recall Details · AllClear