IMMULITE and IMMULITE 1000 Total Testosterone Assay- In vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers for the quantitative measurement of total testosterone in serum and heparinized plasma.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Confirmed an average positive bias of 40% for testosterone concentrations of approximately 660 ng/dL (22.9 nmol/L) and above with the IMMULITE/IMMULITE 1000 Total Testosterone Assay when compared to the IMMULITE 2000/IMMULITE 2000 XPi Total Testosterone Assay. Two (2) lots of Total Testosterone assays for the IMMULITE/IMMULITE 1000, kit lots #515 and #516 were actively distributed to customers when the issue was discovered by Siemens, and they are subject to the correction or removal in this recall (867 total quantity of reagents distributed for each lot). The violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. In both males and female, testosterone results are not used in isolation but are interpreted in conjunction with clinical history and symptomology as well as with other laboratory testing (e.g. gonadotropins and other sex steroids). A delay in the diagnosis may be
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAChlorthalidone Tablets, USP, 25 mg, [100 or 1000] Tablets pr bottle, Rx only, Manufactured by: Inventia Healthcare Limited, Additional Ambernath, M.I.D.C., Ambernath (East) - 421506, INDIA. Distributed by: Risiong Pharma Holdings, Inc., East Brunswick, NJ 08816. NDC 100-tablet bottle: 64980-599-0...2026-06-05
- FDASiemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay2026-05-07
- CPSCBayer Recalls 6 mL Size Afrin Original Nasal Spray Bottles Due to Risk of Serious Injury or Illness from Child Poisoning; Violates Mandatory Standard for Child-Resistant Packaging2026-04-30
- FDAdirectCHECK ACT-LR whole blood quality control, level 2 Model/Catalog Number: DCGLR-22026-04-30
- FDABrand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.2026-04-22
- FDABioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.2026-04-22
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;2026-04-02
- FDABiofire Spotfire Respiratory/Sore Throat Panel REF: 4234852026-04-01
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →