FDANovember 1, 2023device

Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 362032 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; 362033 UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; 362034 Priming volume: 19mL, length: 120in, SafeLine; 362035 INFUSOMAT SP...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, which will stop the infusion and interrupt therapy. Interruptions or delays in medication administration may result in serious injury to patients which may be life-threatening (e.g. interruption of a vasopressor to a hemodynamically unstable patient) or may lead to death.

What to do

FDA enforcement status: Ongoing

Brands named

b braun medical

UPCs

040469648378070404696418255604046964837821040469641825700404696469904704046964182631040469555927150404696418267904046964837845040469642941740404696429419804046964294259