FDASeptember 24, 2020device
ZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 28 mm, 10 degree inclined face, O.D. Cup with Spacers, 49 mm Number: 00-8065-646-28
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Sterilant (hydrogen peroxide) used in acetabular cup hip prosthesis component was not evaluated to confirm the biocompatibility of the residual sterilant. Cannot rule out the potential for adverse tissue reactions.
What to do
FDA enforcement status: Terminated
Brands named
zimmer biometzimmer
UPCs
00806564628
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOff-Axis Alliance Glenoid, Right, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 1100404762026-05-20
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 1100404442026-05-20
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 1100402402026-05-20
- FDAOff-Axis Comprehensive Shoulder System, Large, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 1100402422026-05-20
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 1100402412026-05-20
- FDABrand Name: Persona Revision Product Name: Persona Revision¿ Trabecular Metal¿ Femoral Distal Augment Model/Catalog Number: 42-5572-066-10 Software Version: N/A Product Description: Persona Revision¿ Trabecular Metal¿ Femoral Distal Augment Size 9, 9+ 10mm Thicknes Component: N/A2026-05-11
- FDABrand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A2026-03-30
- FDABrand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 5000TS Tourniquet Systems Model/Catalog Number: 60500010100 Software Version: Software version v2.04 or prior are within scope of the recall for A.T.S.¿ 5000TS Product Description: A.T.S 5000TS Tourniquet Systems Component: N/A2025-12-24