Model: CT 5300; Product Code (REF): 728285; Software Versions: 5.1.0.X & 5.1.1.X;
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wro
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAPhilips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.2026-06-05
- FDAGSMS Incorporated, NIACIN EXTENDED-RELEASE TABLETS, USP, 1,000 MG, 90 tablets, Rx only, Manufactured by Kremers Urban Pharmaceuticals Inc., a subsidiary of Lannett, Inc., Seymour, IN 47274, Packaged by GSMS Incorporated, Camarillo, CA 93012. NDC 51407-268-90.2026-05-29
- FDAPhilips Avalon Fetal Monitor, FM 30 Part numberM2703A2026-05-26
- FDAPhilips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A2026-05-26
- FDAPhilips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 722006 for Allura Xper FD20; Model # 722008 for Allura Xper FD20 Biplane; Model # ...2026-05-18
- FDAPhilips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Detector; Model # 708036 for MultiDiagnost Eleva; Model # 708037 for MultiDiagnost Eleva with Flat Detector; Model # 708038 for MultiDiagnost Elev...2026-05-18