FDAOctober 11, 2016device

ARCHITECT Lactic Acid Part # 09D891T21 The Lactic Acid assay is used for the quantitation of lactic acid in human plasma.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Abbott has identified negative interference from the drug N-Acetyl Cysteine (NAC) with the ARCHITECT Lactic Acid reagent (LN 9D89-21).

What to do

FDA enforcement status: Terminated

Brands named

abbott laboratoriesabbott