FDANovember 17, 2015device
Olympus ENDOEYE HD II Video Telescopes. Models WA50040A ENDOEYE HD II, 10 mm, 0 degree, WA50042A ENDOEYE HD II 10 mm, 30 degree, W7100603 Repair cable 10 mm 0/30 degree. The ENDOEYE video Telescopes are used with other supporting equipment for endoscopy and endoscopic surgery within the thoracic ...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
A damaged temperature sensor at the ENDOEYE tip caused the distal end to become abnormally hot. Excessive heating of the ENDOEYE distal end could result in patient or user injury.
What to do
FDA enforcement status: Terminated
Brands named
olympus
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2113.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-21122026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2118.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2110. May be included with the Olympus OER-ELITE.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-21112026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2117.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2116.2026-03-25
Own something like this?