Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument Kit, Mode: 9736188, and the Bundle Autoguide Reg Kit.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
A tracker that holds and guides neurosurgical surgical instruments may be subject to separation/weld failure, where the tracker connects to the instrument tube. If a weld separation occurs and a navigation inaccuracy is experienced, it may result in prolonged procedure and tissue injury. Per 11/24/21 firm email: The Bundle Autoguide Reg Kit does not have a model number. The Bundle Autoguide Reg Kit is internal terminology and not visible to the customer and are used as sales and shipping configuration only. When this sales bundle is chosen by an internal Medtronic Sales Representative the devices included in the bundle are individually packaged and labeled and sent to the customer in a single shipment. The devices included in the bundle are ordered, packed, and shipped as individual medical devices with unique model numbers, which is the reason that the Autoguide Tracker (28248) and the Stealth Autoguide Basic Instrument Kit (9736188) are the only part numbers subject to this FCA.
What to do
FDA enforcement status: Ongoing
Brands named
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAMedtronic O-arm O2 Imaging System. Model Number: BI70002000.2026-05-19
- FDAMedtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted, for parkinsonian tremor2026-05-13
- FDAMedtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002; Stimulator, electrical, implanted, for parkinsonian tremor2026-05-13
- FDAOctopus Evolution AS Tissue Stabilizer, Model TS25002026-05-12
- FDAOctopus Evolution Tissue Stabilizer, Model TS20002026-05-12
- FDAOctopus 4 Tissue Stabilizer, Model 294002026-05-12
- FDAMedtronic GUNDRY Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94115T; Cardiopulmonary bypass vascular catheter2026-04-21