FDANovember 3, 2015device

MYLA CLI V3.X TO V4.1 ML350 SERVER Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

MYLA¿ software connected to a Laboratory Information System (LIS) and a VITEK¿ 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID.

What to do

FDA enforcement status: Terminated

Brands named

biomerieux