FDASeptember 25, 2015device

MiniMed 620G, 3.0mL mg/dL, Model No. MMT-1710 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is indicated for...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.

What to do

FDA enforcement status: Terminated

Brands named

medtronic minimedmedtronic