MiniMed 640G, 3.0mL mmol/L, 3.0mL mg/dL, Model No. MMT-1711 and MMT-1712 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the p...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.
What to do
FDA enforcement status: Terminated
Brands named
medtronic minimedmedtronic
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAMedtronic O-arm O2 Imaging System. Model Number: BI70002000.2026-05-19
- FDAMedtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted, for parkinsonian tremor2026-05-13
- FDAMedtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002; Stimulator, electrical, implanted, for parkinsonian tremor2026-05-13
- FDAOctopus Evolution AS Tissue Stabilizer, Model TS25002026-05-12
- FDAOctopus Evolution Tissue Stabilizer, Model TS20002026-05-12
- FDAOctopus 4 Tissue Stabilizer, Model 294002026-05-12
- FDAMedtronic GUNDRY Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94115T; Cardiopulmonary bypass vascular catheter2026-04-21
- FDAMedtronic DLP Retrograde Cannula Manual-Inflate Silicone Body 15FR, Model 94965; Cardiopulmonary bypass vascular catheter2026-04-21
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →