FDASeptember 29, 2020device

v|tome|x L Series non-destructive testing cabinet x-ray systems

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm up" mode was not completed after initializing the system when first turned on.

What to do

FDA enforcement status: Terminated

Brands named

ge inspection technologies