FDAOctober 15, 2021device
Karl Storz Udel Sterilization Tray Instructions for Use, Tray Part Number/Description: 39301ACTS/CAMERA TRAY (STERRAD, INLINE CAMERA HEADS); 39301AS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES 4MM X 18CM); 39301BCTS/CAMERA TRAY (INLINE/C-MOUNT CAMERA HEADS); 39301BS/RIGID TELESCOPE TRAY (STERRA...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
To ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) udel trays may be used with flexible endoscopes with lumen diameters greater-than-or-equal to 1.2mm and maximum length of 845mm and 2) ethylene oxide is excluded as a sterilization modality.
What to do
FDA enforcement status: Ongoing
Brands named
karl storz endoscopykarlkarl storz
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDATangent Endoscopy, Tangent Single Use Digital Catheter. Model Numbers: (1) TNG4002-IND. (2) TNG-4007-6pk.2026-03-23
- FDAKARL STORZ - ENDOSKOPE, REF: 27000L,Uretero-Renoscope, 7 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 01232024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27011K, Ureteroscope, 8 Fr., 6¿, 34cm, CE 01232024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27003L, MICHEL Uretero-Renoscope, 9.5 Fr., 43cm, RxONLY, CE01232024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27030KA, Cysto-Urethroscope, NON STERILE, CE2024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27002KP, Pediatric Uretero-Renoscope 7.3 Fr., NON STERILE, CE 01232024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27001L, Uretero-Renoscope, 8 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 01232024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27002L, Uretero-Renoscope, 9.5 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 01232024-12-19