Atellica CH 930 Analyzer with Software version: V1.25.1 and lower- An automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11067000
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
(1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay Correlation factor; (2) On Board Stability (OBS) Not Updating with Manual Change for Open Channel Assays may continue to use the reagent past its expiration date leading to potential erroneous patient sample results
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAChlorthalidone Tablets, USP, 25 mg, [100 or 1000] Tablets pr bottle, Rx only, Manufactured by: Inventia Healthcare Limited, Additional Ambernath, M.I.D.C., Ambernath (East) - 421506, INDIA. Distributed by: Risiong Pharma Holdings, Inc., East Brunswick, NJ 08816. NDC 100-tablet bottle: 64980-599-0...2026-06-05
- FDASiemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay2026-05-07
- CPSCBayer Recalls 6 mL Size Afrin Original Nasal Spray Bottles Due to Risk of Serious Injury or Illness from Child Poisoning; Violates Mandatory Standard for Child-Resistant Packaging2026-04-30
- FDAdirectCHECK ACT-LR whole blood quality control, level 2 Model/Catalog Number: DCGLR-22026-04-30
- FDABrand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.2026-04-22
- FDABioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.2026-04-22
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;2026-04-02
- FDABiofire Spotfire Respiratory/Sore Throat Panel REF: 4234852026-04-01
Own something like this?