FDANovember 12, 2015device

Alaris EtCO2 Module, Model 8300. A capnograph that continuously monitors end tidal carbon dioxide (EtCO2), fractional inspired carbon dioxide (FiCO2), and respiratory rate (RR).

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service using a higher than specified concentration of CO2. The effect to the EtCO2 module tested with a higher concentration of CO2 could lead to an incorrect measurement of CO2.

What to do

FDA enforcement status: Terminated

Brands named

carefusion 303carefusion