FDAOctober 13, 2020device

Verigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP, IVD - Product Usage: IVD sample processor, which utilizes a single-use disposable test consumable and a self-contained Verigene Test Cartridge f...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENESP devices.

What to do

FDA enforcement status: Terminated

Brands named

luminex

UPCs

00857573006010

Recall history