Leica MICROSYSTEMS Provido, Part Numbers 10448950 (Premium), 10448976 (Standard). Surgical microscope system

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.

What to do

FDA enforcement status: Ongoing

Brands named

leica microsystemsleica

Recall history