FDAOctober 13, 2020device
Verigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay Luminex, VERIGENE Gram-Negative Blood Culture Nucleic Acid Test, BC-GN, IVD - Product Usage: IVD sample processor, which utilizes a single-use disposable test consumable and a self-contained Verigene Test Cartridge f...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENE SP devices.
What to do
FDA enforcement status: Terminated
Brands named
luminex
UPCs
00857573006010
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAVERIGENE Gram Negative Blood Culture (BC-GN) Nucleic Acid Test, REF 20-006-021; for use with the sample-to-results VERIGENE System; invitro diagnostic test2025-04-16
- FDANxTAG Respiratory Pathogen Panel, REF: I051C04472024-09-05
- FDANxTAG Respiratory Pathogen Panel +SARS-CoV-2, REF: I056C04682024-09-05
- FDAVerigene Gram Positive Blood Culture (BC-GP) Nucleic Acid Test2024-06-20
- FDALuminex VERIGENE BC-GP Utility Tray (5-Pack Carrier), Gram-Positive Blood Culture Nucleic Acid Test, REF 20-011-018, a component of the Luminex VERIGENE BC-GP Utility Kit, REF 20-012-018, containing 20 tests.2023-11-07
- FDAVerigene Enteric Pathogens Nucleic Acid Test (EP), REF 30-002-0232023-06-05
- FDAVERIGENE Clostridium difficile Nucleic Acid Test, REF 30-002-0222023-06-05
- FDAARIES SARS-CoV-2 Assay, REF: 50-100472022-09-14
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →