FDANovember 17, 2015device

The Fuhrman Pleural/Pneumopericardial Drainage Set The Fuhrman Pleural/Pneumopericardial Drainage Set consists of pigtail catheter, access needle, dilator, and wire guide. The set also includes a multipurpose tubing adapter, and three-way stopcock.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Stryker Sustainability Solutions is recalling the Fuhrman Pleural & Pneumopericardial Drainage Set due to receiving complaints that it broke off in the pleural cavity while inserting the device into the patient.

What to do

FDA enforcement status: Terminated

Brands named

stryker sustainability solutionsstrykerstryker sustainability