Minicon Clinical Sample Concentrators for In Vitro Diagnostic Use 9031 (Model: Minicon B-15) & 9051 (Model: Minicon CS-15) Minicon concentrators are non-sterile, disposable, multiwall ultrafiltration devices. Minicon¿ clinical sample concentrators are for in vitro diagnostic use and are intended ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

In vitro diagnostic devices, used greater than 6 months after date of manufacture, may have increased concentration times of greater than the specification of 3 hours and up to 24 hours, which may lead to delays in carrying out the subsequent analysis tests.

What to do

FDA enforcement status: Terminated