FDAOctober 2, 2023device

STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB Meter as an in vitro diagnostic for blood ketone values.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis.

What to do

FDA enforcement status: Ongoing

Brands named

stanbio laboratory lpstanbiostanbio laboratory