FDAOctober 3, 2025device

MC3 VitalFlow Console, REF 58100; Blood pump of ecmo

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.

What to do

FDA enforcement status: Ongoing

Brands named

medtronic perfusion systemsmedtronicmedtronic perfusion

UPCs

10850007732247

Recall history