FDADecember 15, 2017device

Precedence 6 Model # 882351 Product Usage: Precedence (cleared as Griffin) is an imaging system combining the acquisition of single photon nuclear medicine images and images from an x-ray computed tomography system. Precedence may produce non-attenuation corrected and attenuation corrected images...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.

What to do

FDA enforcement status: Terminated

Brands named

philips medical systems clevelandphilipsphilips medical

UPCs

300070200213000703002030007010017300070300213000706002430007070025300070500233000705002230007030016

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →