Precedence 16 Model # 882350 Product Usage: Precedence (cleared as Griffin) is an imaging system combining the acquisition of single photon nuclear medicine images and images from an x-ray computed tomography system. Precedence may produce non-attenuation corrected and attenuation corrected image...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.

What to do

FDA enforcement status: Terminated

Brands named

philips medical systems clevelandphilipsphilips medical