FDASeptember 30, 2025device

MEDLINE convenience kits labeled as: 1) MINOR SINGLE BASIN PACK, REF DYNJ37698L; 2) BASIN PACK, REF DYNJ50123D; 3) MAJOR BASIN PACK, REF DYNJ59816A; 4) OSC BASIN PACK, REF DYNJ68217A.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

What to do

FDA enforcement status: Ongoing

Brands named

medline industries lpmedlinemedline industries

UPCs

1019532743552340195327435524101953272019994019532720199010195327331573401953273315741019532735252340195327352524