FDASeptember 30, 2025device

MEDLINE convenience kits labeled as: 1) GENERAL LAPAROSCOPY, REF CDS930027W; 2) GENERAL MAJOR CDS, REF CDS980245X; 3) BASIC GENERAL CDS, REF CDS980566F; 4) TOTAL KNEE CDS, REF CDS982823R; 5) BASIC CDS, REF CDS982936U; 6) UPPER EXTREMITY CDS, REF CDS983907M; 7) MAJOR ABDOMINAL CDS, REF CDS983908Q;...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

What to do

FDA enforcement status: Ongoing

Brands named

medline industries lpmedlinemedline industries

UPCs

101953276014924019532760149310193489848434401934898484351019348944544240193489445443101953275844814019532758448210195327185220401953271852211019532758449840195327584499