FDAOctober 28, 2024device

smiths medical PORTEX Tracheal Tube Exchange Guide, 10Ch x 70cm, REF 14-504-76

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.

What to do

FDA enforcement status: Ongoing

Brands named

smiths medical asdsmithssmiths medical

UPCs

35021239002593

Recall history