CAS-930A Flat Panel Monitor (FPD) used in conjunction with the Canon Interventional Angiography System, Model numbers: Infinix-8000C and Infinix-8000H - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The sy...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage to the patients blood vessels due to removal without visibility.

What to do

FDA enforcement status: Terminated

Brands named

canon medical systemcanoncanon medical