FDAOctober 10, 2023device

VITROS 5600 Integrated System Product Code 6802413 and Refurbished Product Code 6802915 Running Software Version 3.8.0

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.

What to do

FDA enforcement status: Ongoing

Brands named

ortho clinical diagnosticsorthoortho clinical

UPCs

1075875000274010758750007110

Recall history