FDAJune 5, 2017device
Zimmer Natural Nail System Tear Drop Guide Wire 3.0 mm Diameter, 70 cm length, Item Number 47-2490-098-00 Product Usage: The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wir...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.
What to do
FDA enforcement status: Terminated
Brands named
zimmer biometzimmer
UPCs
47249009800
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOff-Axis Alliance Glenoid, Right, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 1100404762026-05-20
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 1100404442026-05-20
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 1100402402026-05-20
- FDAOff-Axis Comprehensive Shoulder System, Large, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 1100402422026-05-20
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 1100402412026-05-20
- FDABrand Name: Persona Revision Product Name: Persona Revision¿ Trabecular Metal¿ Femoral Distal Augment Model/Catalog Number: 42-5572-066-10 Software Version: N/A Product Description: Persona Revision¿ Trabecular Metal¿ Femoral Distal Augment Size 9, 9+ 10mm Thicknes Component: N/A2026-05-11
- FDABrand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 00-5049-011-00 Software Version: N/A Product Description: Mixing Bowl and Spatula Component: N/A2026-03-30
- FDABrand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 5000TS Tourniquet Systems Model/Catalog Number: 60500010100 Software Version: Software version v2.04 or prior are within scope of the recall for A.T.S.¿ 5000TS Product Description: A.T.S 5000TS Tourniquet Systems Component: N/A2025-12-24
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →