FDAOctober 9, 2024device

Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.

What to do

FDA enforcement status: Ongoing

Brands named

shimadzu corp analytical measuring instruments divshimadzushimadzu corp

UPCs

045402174201310454021742012404540217420155

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.