FDAOctober 9, 2024device

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.

What to do

FDA enforcement status: Ongoing

Brands named

shimadzu corp analytical measuring instruments divshimadzushimadzu corp

UPCs

0454021742058204540217420575