FDANovember 18, 2020device

Trinias, Digital Angiography System containing a MH-200, Ceiling Suspended C-arm Support - Product Usage: intended to be used for cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiogr...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The firm has identified a problem with the control software for the celling arm of the X-Ray System . This software issue could cause the ceiling arm of the device to move after the operator has released the movement lever. This could result in possible collision or harm to the patient or system operator.

What to do

FDA enforcement status: Ongoing

Brands named

shimadzu medical systemsshimadzushimadzu medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.