Philips Allura systems with monoplane fixed ceiling mounts: 722001 Allura Xper FD10C 722003 Allura Xper FD10 722006 Allura Xper FD20 722010 Allura Xper FD10 722012 Allura Xper FD20 722015 Allura Xper FD20 OR Table 722016 Integris H5000 C / Allura 9C 722018 Integris Allura 9 722022 Allura Xper FD1...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose

What to do

FDA enforcement status: Ongoing