FDANovember 17, 2015device

Valve XS Atrium Retractor FC429R 52154756 The Valve XS Atrium Lift Retractor is used in heart surgery. The retractor is used for the temporary dilation of the left atrium in mitral valve surgery and for assistance in tricuspid valve surgery.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

AAG received a customer complaint that corrosion had been found by a hospitals central sterile service department on the ball joint of an FC424R blade. Investigation revealed that the corrosion was due to incorrect material used in the manufacturing of the disk spring of the ball joint by the supplier

What to do

FDA enforcement status: Terminated

Brands named

aesculap

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Valve XS Atrium Retractor FC429R 52154756 The Valve XS Atrium Lift Retractor is used in heart surgery. The retractor is used for the temporary dilation of the left atrium in mitral valve surgery and for assistance in tricuspid valve surgery. — Recall Details · AllClear