FDAOctober 24, 2019device

NC Emerge Monorail PTCA Dilatation Catheter. Sterilized using ethylene oxide. Packaged in the following sizes: 1. REF/Catalog No. H7493926708350, NC EMERGE MR US 3.50MM X 8MM; 2. REF/Catalog No. H7493927608350, NC EMERGE MR OUS 3.50MM X 8MM; 3. REF/Catalog No. H7493927608400, NC EMERGE MR OUS 4.0...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Certain dilation and extension catheters are being recalled because these catheters may have missed the hydrophilic coating step, which creates a lubricious surface on the devices, during the manufacturing process.

What to do

FDA enforcement status: Terminated

Brands named

boston scientificboston

UPCs

871472984628487147298470078714729847021871472984707687147298463768714729847090871472984638387147298464068714729847137871472984643787147298471758714729847236

Recall history