FDAOctober 22, 2024device

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.

What to do

FDA enforcement status: Ongoing

Brands named

micro xmicro

UPCs

9357123000037