Behind-the-Ear hearing aids, Model Numbers: Q962-DRW, T561-DRW, T562-DRW, T762-DRW, T777-DW, AMZ964-DRW, ET588-DWH, ET598-DW, ET788-DWH, ET798-DW, ET988-DWH, ET998-DW, FT861-DRW, FT862-DRW, RE561-DRWC, RE762-DRW, RE961-DRWC, RE962-DRW, PZ862-DRW, LT561-DRW, LT561-DRWZ, LT562-DRW, LT588-DW, LT761-...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
GN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported.
What to do
FDA enforcement status: Terminated
Brands named
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- CPSCHearing Lab Technology/Lucid Audio Recalls Children's HearMuffs Due to Burn and Injury Hazards from Rupturing Alkaline Batteries2022-09-08
- FDAIn-the-Ear hearing aids, Model Numbers: T9ITC-DW-LP, FT6ITC-D-MP, FT6ITE-D-MP, FT8CIC-MP, FT8ITC-W-HP, FT8ITC-W-LP, FT8ITC-W-MP, FT8ITE-DW-HP, FT8MIH-W-HP, FT8MIH-W-MP, LT5CIC-MP, LT5CIC-UP, LT5IIC-LP, LT5ITC-DW-MP, LT5ITC-DW-UP, LT5ITE-DW-HP, LT5ITE-DW-MP, LT5ITE-DW-UP, LT5MIH-W-HP, LT5MIH-W-MP,...2019-09-18
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →