FDANovember 18, 2015device

Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Potential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout, the 4D tabcard will show an incorrect, too small length measurement. This could result in selecting a device of the wrong size, which then needs to be exchanged.

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical