epoc NXS Host Blood Analysis System, Siemens Material Numbers 11413497 (EU), 11413475 (US), 11413506 (CA), 11413517 (MX), 11413518 (ROW), 11413498 (JP), 11413583 (CN, KR), 11413879 (IN) with software versions from 4.9.4 to the current version 4.11.11

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Siemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under certain conditions, a software problem is causing erroneous results to be printed. When the results are run on the epoc NXS Host and Delete Blood Tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test. The worst possible outcome may result in a failure to diagnose hyperkalemia, due to an erroneous depressed potassium level or inappropriate treatment to hyperglycemia due to erroneous glucose reading.

What to do

FDA enforcement status: Ongoing

Brands named

siemens healthcare diagnosticssiemenssiemens healthcare

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
epoc NXS Host Blood Analysis System, Siemens Material Numbers 11413497 (EU), 11413475 (US), 11413506 (CA), 11413517 (MX), 11413518 (ROW), 11413498 (JP), 11413583 (CN, KR), 11413879 (IN) with software versions from 4.9.4 to the current version 4.11.11 — Recall Details · AllClear