FDAOctober 21, 2019device

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

What to do

FDA enforcement status: Terminated

Brands named

integra lifesciencesintegra