FDAMay 30, 2023device

Alinity ci-series System Control Module, REF: 03R70-01

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3) Customer requests mechanism to prevent user error when creating new lots of value assigned Calibrators; 4) Error in processing tests after CCCintController encountered an error with the assay information downloaded from SCC; 5) The assay editor does not correctly check sample dilution total volume limits; 6) User-Applied Labels not adhering to Reagent Bottles; 7) ICT Module did not expire after warranty was exceeded. These performance issues could lead to erroneous results for multiple analytes.

What to do

FDA enforcement status: Ongoing

Brands named

abbott laboratoriesabbott

UPCs

00380740137410

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.