FDADecember 14, 2015device

FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.

What to do

FDA enforcement status: Terminated

Brands named

becton dickinson and company bd biosciencesbectonbecton dickinson